- Research
- Open access
- Published:
A1 pulley cutting potential and safety of three ultrasound-guided percutaneous A1 pulley release techniques for trigger finger: a cadaveric study
Journal of Orthopaedic Surgery and Research volume 20, Article number: 215 (2025)
Abstract
Background
There is no consensus as to the best technique for percutaneous trigger finger release.
Methods
This assessor-blinded study compared three ultrasound-guided percutaneous trigger finger release techniques using a needle (N), a needle-knife (NK), and a specially designed knife (K). Three physicians simulated A1 pulley release surgery on 56 fingers of 14 fresh-frozen hand cadaver body donors. Both the physicians and the fingers included were randomly selected.
Results
The results of repeated-measures ANOVA revealed significantly longer cuts for the NK and K techniques, than for the N technique, both absolute (mean ± SD) (NK = 5.55 ± 3.07 mm, K = 6.29 ± 4.07 mm, and N = 2.02 ± 3.46 mm; N vs. NK p = 0.015, N vs. K p = 0.002, and NK vs. K p = 1.000), and cut percentage in relation to the total pulley length (NK = 51.61 ± 28.34%, K = 54.63 ± 33.72% and N = 18.24 ± 31.09%; N vs. NK p = 0.008, N vs. K p = 0.003, and NK vs. K p = 1.000). No neurovascular bundle injuries were found upon dissection. The overall complication rate was 11%, with no significant differences among the three techniques. Only one major tendon injury occurred in the NK group.
Conclusions
In this cadaveric study, the NK and K techniques were more effective at releasing the A1 pulley than the N technique. All three techniques have emerged as equally safe.
Background
The trigger finger (TF) is a common condition that affects up to 3% of the general population [1]. Open TF release is among the most frequently performed interventions on the hand, with successful outcomes in a high percentage of cases [2,3,4]. However, a number of complications may be given when open surgery is performed [5]. To improve treatment efficiency, and reduce recovery time, infection risk and the costs of open surgery, numerous percutaneous TF release techniques have been described [3, 6,7,8,9,10,11,12,13,14,15,16].
Percutaneous blind release of TF has a number of complications that should not be underestimated [17] so endoscopic techniques [17] or ultrasound (US)-guided procedures [6,7,8, 18] have been described without evidence yet on which is best. A number of recent systematic reviews and metanalysis found US-guided percutaneous A1 pulley release to be advantageous over open surgical release in terms of improved QDASH score in the first month, accounting for a quicker recovery from surgery, quicker return to normal activities, and an earlier discontinuation of oral analgesia [18, 19]. Long term conclusions such as recurrence rates, have been brought up in favor of US-guided techniques but still lack solid evidence [18, 20].
When using an US-guided technique, a wide range of cutting instruments, including conventional or modified needles, needle-knives and special knives, have been tested [6, 7, 9,10,11,12,13, 21,22,23,24]. There are also many uncertainties about the efficacy of pulley release and the safety of US-guided percutaneous approaches, especially because of the potential risk of tendon and/or neurovascular injury, as reported outcomes vary widely [14].
The main aim of this cadaveric study was to compare the cutting potential and safety, in terms of structural injury to surrounding neurovascular or tendinous structures, of US-guided percutaneous A1 pulley release via a needle, a needle knife, and a jawed knife specially designed for pulley release (Fig. 1).
Methods
Design
Three surgical techniques were performed on 56 triphalangeal fingers from 14 cryopreserved body donor hands (9 females, 8 right), previously left to thaw at normal room temperature. Then they underwent further dissection. The anatomy and disposition of the thumb pulley system is much different from that of triphalangeal fingers and has some safety particularities that require different A1 pulley release techniques from those performed on triphalangeal fingers. For this reason, thumbs were not included in the present study. As no additional tissue preservation technique was used, cadaver specimen tissue consistency was preserved. Data were collected in two sessions, one in November and one in December 2022. Institutional review board approval was obtained prior to the study. The body donation program and use of cadaver tissue were in compliance with the current national legislation regarding ethics in research. None of the cadaver hands showed signs of trauma, deformities or surgical scars.
The cadaveric specimens were randomly distributed among three physicians with extensive experience (over 20 years) of US imaging and US-guided interventional procedures but without experience in the surgical techniques to be performed. This way we sought to analyze reproducibility of techniques already described in the indexed literature. Fingers were assigned to the three physicians via a code to indicate the hand and finger number. A statistician not involved in the study assigned a computer-generated randomization sequence to each physician, stratified by the finger and the technique. The physicians were unaware of the randomization sequence, which was concealed until the time of the simulated surgery. Another investigator, blinded to the technique and physician, performed anatomical dissection of the fingers after the procedure to assess surgical outcomes and complications.
Procedures and surgical techniques
Cadaveric limbs were placed in a supine position with the hands facing upwards on the table. All three procedures started with a US examination along the flexor tendon axes from proximal to distal. The same US device (APLIO i800; Canon Medical Systems, Tokyo, Japan) with a high-frequency hockey stick probe (18 MHz) was used for all US scans by all the physicians. In all the cases, the longitudinal axis of the tendon, and the proximal and distal edges of the A1 pulley were labeled on the skin via the method described by Rojo-Manaute et al. [25]. These authors used a needle positioned between the transducer and skin to trace the long axis of the A1 pulley (Fig. 2a). The proximal edge of the pulley is defined by the metacarpal head-neck junction, and the distal edge is defined by the junction between the base and shaft of the proximal phalanx [25].
Skin markings prior to the procedure via an ultrasound probe. (a) Longitudinal line of the flexor tendon. (b) Marking the distal and proximal margins of the A1 pulley. (c) Cross-sectional ultrasound image showing the needle shadow. (d) Longitudinal ultrasound image showing the needle shadow to mark the proximal and distal limits of the A1 pulley. The arrow indicate the needle reverberation. The asterisk indicate the flexor tendons. MH: metacarpal head. PPh: proximal phalanx
Once the marks were drawn on the skin (Fig. 2b), and with US guidance, the flexor tendons were visualized in the cross-sectional and longitudinal axes (Fig. 2c and d), the three physicians performed one of the three following techniques:
-
(1)
Needle technique. This technique uses a 21 G x 50 mm needle with the tip curved at 140º and the bevel positioned laterally as a cutting instrument [10]. The US probe was placed on the long axis over the flexor tendon at the level of the metacarpophalangeal joint. The needle was introduced into the distal third of the proximal phalanx (Fig. 3a). Then, under US guidance, using both the long and short axis view, longitudinal incisions were made from distal to proximal, attempting to cut the pulley in ten similar moves (Fig. 3b).
Needle technique. (a) Inserting a 21 G needle under ultrasound guidance with a stick probe. (b) Ultrasound image identifying the needle penetrating the deepest most distal part of the A1 pulley. MH: metacarpal head. PPh: proximal phalanx. The arrowheads indicate the A1 pulley. The arrow indicates the needle. The asterisks indicate the flexor tendons
-
(2)
Needle-knife technique. This technique is based on an approach perpendicular to the skin surface, which generates a greater cutting force on the pulley. In contrast to the original description by Eastwood [15], a Nokor® (Becton Dickinson and Co., Franklin Lakes, New Jersey, USA) 18 G needle with a triangular scalpel blade at its tip was used, allowing for fewer cutting movements in a more linear way than with the beveled needle. After performing the corresponding A1 pulley skin markings, the needle-knife was introduced perpendicular to the palmar skin surface in the middle of the A1 pulley along the longitudinal axis of the flexor tendon. Passive flexion and extension movements of the finger were used to ensure that the needle was not in contact with the flexor tendon. Six cutting movements were performed from distal to proximal with the needle-knife.
-
(3)
Knife technique. For this technique, a specially designed jaw-bladed knife was used to section the A1 pulley (Kemis H1, Newclip Technics, Haute Goulaine, France). This cutting instrument has a 3 mm wide blade consisting of a shorter upper and longer lower prong (Fig. 4a). A 2 mm skin incision was made 1 cm proximal to the proximal edge of the A1 pulley. After expanding the space under the A1 pulley via curved mosquito artery forceps (Fig. 4b), the knife was introduced with the longer lower jaw under the pulley and over the tendon to cut the pulley under direct US guidance (Fig. 4c and d). A single cutting move was performed.
Knife technique. (a) Close up of the tip of the knife showing the long and short prongs. (b) Enlarging the space via mosquito artery forceps. (c and d) The knife was introduced under ultrasound guidance to perform the release. The arrowheads indicate the shorter upper jaw of the knife. The arrow indicates the longer lower jaw of the knife
Anatomic dissection and outcomes
After the surgical procedure, an expert anatomist blinded to the technique performed the dissection of the samples and recorded the following outcomes for each digit: (1) complete release of the A1 pulley achieved or not, (2) digital nerve or vascular injury, (3) tendon injury noted as minor (minimal surface abrasion or excoriation affecting less than 10% of the tendon thickness) or major (visible laceration affecting more than 10% of the tendon thickness or potentially clinically relevant injury), (4) A2 pulley injury, (5) minimum transverse distance from the pulley incision to the digital collateral nerve, and (6) A1 pulley incision site determined by inspection during the anatomic dissection (midline, ulnar, radial, or oblique). The length of the incision on the A1 pulley was measured via a precision caliper (Qfun Ò digital caliper, 0–150 mm, CN, China) (Fig. 5). To assess the degree of A1 pulley division, the longitudinal length of the A1 pulley divided by the total length of the A1 pulley was compared and calculated as a percentage. A cut that accounted for the total length of the A1 pulley was considered a complete pulley release. A cut greater than 50% of the total length was considered substantial pulley release. Pulleys with no cuts were recorded as “intact”.
Statistical analysis
Data was analyzed using Stata SE version 18 for Windows (Stata Corp LLC, TX; USA). The absolute and relative frequencies are described as percentages, and continuous variables are described as the means and standard deviations. The chi-square test was used to compare categorical variables, and Fisher’s exact F test was used when the value was less than 5 in at least one of the cells of the n x m contingency tables. Quantitative variables were analyzed via ANOVA to compare the mean data for each technique, with Bonferroni correction. Kruskal-Wallis tests were employed when appropriate. Significance was set at p < 0.05.
Results
The characteristics of the cadavers and fingers treated in this study are summarized in Table 1.
The length of the A1 pulley divided was significantly greater in the needle-knife and knife groups than in the needle group, both in absolute value, and in percentage value (Table 2). Although no significant differences were found between the techniques in terms of the number of complete releases, the needle-knife and knife techniques yielded greater numbers of substantially released pulleys and lower numbers of intact pulleys (Table 2). Although the differences were not statistically significant, there were more tendon and A2 pulley injuries in the NK group (Table 2). The minimum transverse distance between the pulley incision and the closest digital collateral nerve did not differ across the different techniques (Table 2).
The site of pulley release could be determined in 41 of the fingers. No significant differences were found in this variable among the three techniques (Table 3).
Discussion
In this assessor-blinded cadaveric study, our main finding was that when performing an A1 pulley release with US guidance or assistance, it was possible to achieve longer release of the A1 pulley via the needle-knife or knife technique than via the needle technique, whereas the number of structural injuries to surrounding neurovascular or tendinous structures was not statistically significant amongst techniques.
The needle technique has the lesser cutting potential. The aim is to progressively debilitate the A1 pulley with repetitive punctures until it is progressively torn. In the present study, we achieved A1 pulley cuts that were approximately one-third of the length obtained with the other two techniques. The high number of incomplete A1 pulley sections is in agreement with that obtained in previous cadaveric studies using 21G needles [10], which means that the technique is reproducible even for practitioners who are not familiar with it. Although the results in cadaveric samples might not seem very encouraging, in the two-part study by Lapègue et al. [10], TF symptoms were resolved in 82% of patients on the same day of surgery, and the percentage increased to 97% after a 6-month follow-up with additional steroid injections [10]. Thus, adjunct procedures such as US-guided corticoid injections should be considered when needle techniques are used.
As the degree of pulley release accounts for the percentage of cases that cannot be clinically solved by the needle technique, efforts have been made to develop techniques that have greater cutting potential and can achieve greater pulley sections, while maintaining a good safety profile. Adding a small knife to the tip of a needle to enable direct cutting motion instead of a repetitive puncture approach, can be achieved with modified needle-knives with excellent clinical results under US guidance [7]. Seemingly, in the present study, significantly greater sections were obtained compared with the needle technique, and no significant differences in terms of complications were observed.
Knife-based techniques have the greatest cutting potential and have obtained very good clinical and functional results in previous reports, with success rates in the treatment of TF between 94% and 100% [24], and a 90 to 98% rate of A1 pulley complete release in fresh frozen cadaver hands [16, 26]. However, injuries to the A2 pulley account for as many as 12% of the injuries to cadaver fingers [16]. To minimize A2 pulley sections and other potential complications, we used the Kemis H1 knife under US guidance. There were far fewer complete A1 pulley sections, and most of the A1 pulleys were sectioned for more than 50% of the total pulley length, with only one (1.79%) A2 pulley damaged. Further studies should corroborate whether percutaneous US-guided knife techniques achieve the best balance between the length of the pulley cut, the safety of the procedure and the clinical resolution of symptoms as well as how aggressively a surgeon must maintain this balance.
The most frequent complications described in cadaveric studies following percutaneous TF release are longitudinal injuries of the flexor tendons and partial digital artery lacerations [6, 27]. Our data confirm that percutaneous TF release is a safe procedure for the techniques here described. By marking the skin under US guidance, the risk of neurovascular bundle or A2 pulley lesions can be minimized regardless of the cutting instrument used, in contrast to percutaneous A1 pulley release without US guidance, which can lead to potential damage to the flexor tendons or adjacent neurovascular structures [7]. We had no neurovascular injuries in our study, three minor tendon injuries, one major tendon injury and two A2 pulley lacerations affecting less than 25% of the total pulley length. Although the difference was not significant, it should be noted that tendon injuries were three times more common with the needle-knife technique than with the knife technique. This percentage of tendon injuries is much lower than the 65% reported by Calleja et al. [28]. The closest transverse distance from the pulley cut to the neurovascular bundle highlighted the sufficient safety margin of all three techniques.
Further studies should aim to identify the most effective percutaneous A1 pulley sectioning technique by means of a well-designed randomized controlled clinical trial. However, only a cadaver study can determine the specific percentage of A1 pulley sections associated with a specific technique, thus, we believe that the results of the present study can contribute to the design of further research. In this way, the question is raised whether the original needle technique by Lapégue [10] could be compared in a cadaver lab setting to corroborate previous findings describing a similar technique with a thicker 14 G angiocath needle but with a greater achievement of complete releases [29]. However, with the results hereby obtained we favor more the needle-knife or knife techniques.
Additionally, the clinical significance of incomplete A1 pulley release requires further study with longer follow-up, considering recurrences and digit range of motion. In this way number of cutting movements should be analyzed in a more practical way other than standardizing a limited number as done in this study for the process of research methodology.
Finally, as this study was performed in cadaver specimens we could closely assess safety parameters between techniques, in terms of structural injury to nearby structures, and even though we did not obtain significant differences among them, we observed a tendency for the needle-knife technique (which is the only technique not performed under direct US guidance but under US assistance) to have more complications. Increasing the sample size could help to further determine if this difference remains statistically not significant.
Conclusions
The findings of this cadaveric study indicate that TF release via the US-assisted needle-knife or US-guided knife techniques hereby studied, can achieve greater A1 pulley release than the US-guided needle technique. The 21G US-guided needle technique here used, proved to be inefficient at achieving A1 cutting lengths greater than 50% of the A1 total pulley. A good safety profile was obtained in terms of structural injury to nearby neurovascular or tendinous structures, without significant differences in between techniques.
Data availability
No datasets were generated or analysed during the current study.
Abbreviations
- A1:
-
First annular
- A2:
-
Second annular
- K:
-
Knife technique
- N:
-
Needle technique
- NK:
-
Needle-knife technique
- TF:
-
Trigger finger
- US:
-
Ultrasound
References
Baumgarten KM, Gerlach D, Boyer MI. Corticosteroid injection in Diabetic patients with trigger finger: a prospective, randomized, controlled double-blinded study. J Bone Joint Surgery-American Volume. 2007;89:2604–11.
Bonnici AV, Spencer JD. A survey of ‘Trigger finger’ in adults. J Hand Surg. 1988;13:202–3.
Gilberts ECAM, Beekman WH, Stevens HJPD, Wereldsma JCJ. Prospective randomized trial of open versus percutaneous surgery for trigger digits. J Hand Surg. 2001;26:497–500.
Turowski GA, Zdankiewicz PD, Thomson JG. The results of surgical treatment of trigger finger. J Hand Surg. 1997;22:145–9.
Bruijnzeel H, Neuhaus V, Fostvedt S, Jupiter JB, Mudgal CS, Ring DC. Adverse events of Open A1 Pulley Release for idiopathic trigger finger. J Hand Surg. 2012;37:1650–6.
Hoang D, Lin AC, Essilfie A, Minneti M, Kuschner S, Carey J, et al. Evaluation of Percutaneous First Annular Pulley Release: efficacy and complications in a perfused cadaveric study. J Hand Surg [Am]. 2016;41:e165–73.
Pan M, Sheng S, Fan Z, Lu H, Yang H, Yan F et al. Ultrasound-guided percutaneous release of A1 pulley by using a needle knife: A prospective study of 41 cases. Front Pharmacol. 2019;10 MAR:267.
Wang J, Zhao J-G, Liang C-C, Percutaneous Release. Open surgery, or corticosteroid injection, which is the best treatment method for trigger digits? Clin Orthop Relat Res. 2013;471:1879–86.
Jou IM, Chern TC. Sonographically assisted Percutaneous Release of the A1 Pulley: a New Surgical technique for treating trigger digit. J Hand Surg. 2006;31:191–9.
Lapègue F, André A, Meyrignac O, Pasquier-Bernachot E, Dupré P, Brun C, et al. US-guided Percutaneous Release of the trigger finger by using a 21-gauge needle: a prospective study of 60 cases. Radiology. 2016;280:493–9.
Rajeswaran G, Healy J, Lee J. Percutaneous release procedures: trigger finger and carpal tunnel. Semin Musculoskelet Radiol. 2016;20:432–40.
Yang J, Ma B, Zhong H, Zhang Y, Zhu J, Ni Y. Ultrasound-guided percutaneous A1 Pulley Release by Acupotomy (Needle-Knife): a cadaveric study of Safety and Efficacy. JPR. 2022;15:413–22.
Rojo-Manaute JM, Capa-Grasa A, Del Cerro-Gutierrez M, Martinez MV, Chana-Rodriguez F, Martin JV. Sonographically guided intrasheath percutaneous release of the first annular pulley for trigger digits, part 2: randomized comparative study of the economic impact of 3 surgical models. J Ultrasound Med. 2012;31:427–38.
Paulius P KL, Maguina. Ultrasound-assisted percutaneous trigger Finger Release: is it safe? Hand (New York N Y). 2009;4:35–7.
Eastwood DM, Gupta KJ, Johnson DP. Percutaneous release of the trigger finger: an office procedure. J Hand Surg. 1992;17:114–7.
Dunn MJ, Pess GM. Percutaneous trigger Finger Release: a comparison of a new push Knife and a 19-Gauge needle in a cadaveric model. J Hand Surg. 1999;24:860–5.
Pegoli L, Cavalli E, Cortese P, Parolo C, Pajardi G. A COMPARISON OF ENDOSCOPIC AND OPEN TRIGGER FINGER RELEASE. Hand Surg. 2008;13:147–51.
Amro S, Kashbour M, Shaaban Abdelgalil M, Qafesha RM, Eldeeb H. Efficacy of Ultrasound-guided Tendon release for trigger finger compared with open surgery: a systematic review and Meta‐analysis. J Ultrasound Med. 2024;43:657–69.
Garcia HRP, Mund E, Romeiro P. Ultrasound-guided vs. non-guided trigger finger release: a systematic review and meta-analysis. Int Orthop (SICOT). 2024;48:2429–37.
Liang Y, Chen L, Cui Y, Du C, Xu Y, Yin L. Ultrasound-guided acupotomy for trigger finger: a systematic review and meta-analysis. J Orthop Surg Res. 2023;18:678.
Guo D, Guo D, Guo J, McCool LC, Tonkin B. A cadaveric study of the thread trigger Finger Release: the First Annular Pulley Transection through Thread Transecting technique. Hand. 2018;13:170–5.
Hopkins J, Sampson M. Percutaneous tenotomy: development of a novel, percutaneous, ultrasound-guided needle-cutting technique for division of tendons and other connective tissue structures: percutaneous tenotomy. J Med Imaging Radiat Oncol. 2014;58:327–30.
Abdoli A, Asadian M, Banadaky SHS, Sarram R. A cadaveric assessment of percutaneous trigger finger release with 15° stab knife: its effectiveness and complications. J Orthop Surg Res. 2021;16:426.
Nikolaou V, Malahias M, Kaseta M, Sourlas I, Babis G. Comparative clinical study of ultrasound-guided A1 pulley release vs open surgical intervention in the treatment of trigger finger. WJO. 2017;8:163.
Rojo-Manaute JM, Soto VL, De las Heras Sánchez-Heredero J, Del Valle Soto M, Del Cerro-Gutiérez M, Martín JV. Percutaneous intrasheath ultrasonographically guided First Annular Pulley Release: anatomic study of a new technique. J Ultrasound Med. 2010;29:1517–29.
Pages L, Cambon A. Ultrasound-guided percutaneous opening of the A1 pulley with surgical knife on anterograde versus retrograde approach: a comparative cadaver study (40 fingers). Hand Surg Rehabilitation. 2023;42:512–6.
Grinčuk A, Baužys K, Porvaneckas N, Uvarovas V, Rauba G, Ryliškis S. Identification of the location of the A1 pulley combining palpation technique with palm landmarks and percutaneous release of A1 pulley with a 19-gauge needle: a cadaveric study. J Orthop Surg. 2017;25:1–4.
Calleja H, Tanchuling A, Alagar D, Tapia C, Macalalad A. Anatomic outcome of Percutaneous Release among patients with trigger finger. J Hand Surg. 2010;35:1671–4.
Bain GI, Turnbull J, Charles MN, Roth JH, Richards RS. Percutaneous A1 pulley release: a cadaveric study. J Hand Surg. 1995;20:781–4.
Acknowledgements
The authors thank, as part of our research team, José Aramendi for his valuable task in assisting in the design, literature search and data analysis. Also we are thankful to another member of our research team, Xeber Iruretagoinea for his assistance in the literature search and review and data collection during the experimental part. Finally we thank Marimar Ubeda from the Donostia University Hospital Library for her help in the literature search.
Funding
This research received no external funding.
Author information
Authors and Affiliations
Contributions
JF, RB and XS designed the study. FD did the literature search and review. JF, XS, AR, MB, and RB performed the experimental part and the anatomical dissection. FD collected the data. JF and MB analysed the data. MB and FD drafted the first manuscript. JF supervised the manuscript. MB finally revised the manuscript. All authors reviewed the final manuscript versión.
Corresponding author
Ethics declarations
Ethical approval
Ethical approval for this study was obtained from the Institutional Review Board from the University of Barcelona, Faculty of Medicine and Health Sciences, in accordance with current Spanish legislation regarding ethics in cadaveric research.
Consent to participate
Not applicable.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
de la Fuente, J., Blasi, M., Dávila, F. et al. A1 pulley cutting potential and safety of three ultrasound-guided percutaneous A1 pulley release techniques for trigger finger: a cadaveric study. J Orthop Surg Res 20, 215 (2025). https://doi.org/10.1186/s13018-025-05590-w
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s13018-025-05590-w