Purpose: To compare intraocular pressure measured using the Goldmann applanation tonometer with that measured using the ocular response analyzer after congenital cataract surgery.
Bahareh Kheiri +2 more
exaly +2 more sources
Comparison of the new rebound tonometer with Goldmann applanation tonometer in a clinical setting [PDF]
Abstract.Purpose: To evaluate the clinical usefulness of a new rebound tonometer, Icare® PRO (Icare PRO), by comparison with Goldmann applanation tonometry (GAT) in a study on patients with glaucoma.Methods: One hundred and seventy‐two eyes of 86 subjects were enrolled in this study.
Min Kyu Yang +2 more
exaly +3 more sources
Some of the next articles are maybe not open access.
Related searches:
Related searches:
Comparison of the Challenger Digital Applanation Tonometer and the Goldmann Applanation Tonometer
American Journal of Ophthalmology, 1986We compared the Challenger electronic tonometer to the Goldmann applanation tonometer in 70 eyes without corneal abnormalities. There was a good overall correlation between the two machines. Correlations between interobserver readings were excellent, demonstrating that obtaining reproducible measurements is possible with either machine.
J I, Lim, J M, Ruderman
openaire +2 more sources
Comparison of intraocular pressure measured by iCare tonometers and Goldmann applanation tonometer
Canadian Journal of Ophthalmology, 2023To evaluate the relationship between the Goldmann applanation tonometer (GAT) and the iCare PRO and iCare IC200 tonometers in measuring intraocular pressure (IOP) in adult eyes with a diagnosis of glaucoma or glaucoma suspect.One hundred and one eyes from 101 participants diagnosed with glaucoma or glaucoma suspect were evaluated in this study. IOP was
Robin Sachdeva +2 more
openaire +2 more sources
Comparison of the Tono-Pen to the Goldmann Applanation Tonometer
Archives of Ophthalmology, 1988Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry in 142 eyes of 71 patients without corneal problems. Overall we found 63% of the Tono-Pen readings to be within +/- 2 mm Hg of the Goldmann applanation tonometry readings and 77% to be within +/- 3 mm Hg, while 4% were within at least +/- 7 mm Hg difference.
R E, Frenkel, Y J, Hong, D H, Shin
openaire +2 more sources
Background: The aim of this study was to compare the intraocular pressure (IOP) results measured by the iCare rebound tonometer with those obtained by the Goldmann applanation tonometer (GAT) over a wide range of IOP values.
Esther M Hoffmann, N Pfeiffer
exaly +2 more sources
Variability in the Calibration Error of the Goldmann Applanation Tonometer
Journal of Glaucoma, 2011To investigate the variability in the measured calibration error with continued usage of Goldmann applanation tonometers (GATs) having unacceptable calibration error.The study included 132 slit-lamp mounted Goldmann tonometers (Model AT 900 C/M; Haag-Streit, Switzerland).
Nikhil Shreeram, Choudhari +3 more
openaire +2 more sources
Measurement of Goldmann Applanation Tonometer Calibration Error
Ophthalmology, 2009To determine the frequency and possible origins of calibration errors of the Goldmann applanation tonometer and to investigate intraobserver and interobserver agreement in the measurement of such calibration errors.Evaluation of diagnostic test or technology.One hundred thirty-two slit-lamp-mounted Haag-Streit Goldmann tonometers (Model AT 900 C/M ...
Nikhil S, Choudhari +4 more
openaire +2 more sources
Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients
Journal of Glaucoma, 2006To compare the intraocular pressure (IOP) readings taken with the new ICare tonometer and with the Goldmann applanation tonometer (GAT) and to evaluate the influence of central corneal thickness (CCT) on the IOP measurements.One eye of 178 consecutive patients with primary open-angle glaucoma underwent ultrasonic CCT measurement, followed by IOP ...
Paolo, Brusini +4 more
openaire +2 more sources
Comparative Evaluation of Diaton and Goldmann Applanation Tonometers
Ophthalmologica, 2012<b><i>Aim:</i></b> To evaluate the agreement between intra-ocular pressure (IOP) measurements with the Diaton and the Goldmann applanation tonometers (GAT). <b><i>Methods:</i></b> In this cross-sectional observational study, a total of 196 eyes of 196 participants (66 patients with glaucoma, 130 healthy ...
Shveta Jindal, Bali +4 more
openaire +2 more sources