T-wave oversensing during left ventricular pacing causing inappropriate defibrillation. [PDF]
Payne J, Gautam S, Weachter R, Flaker G.
europepmc +5 more sources
Intermittent failure of pacing output caused by algorithm to prevent T-wave oversensing. [PDF]
Oversensing of extracardiac myopotentials can lead to pacemaker inhibition and loss of cardiac output. We describe 2 similar cases in which oversensing of myopotentials and loss of pacemaker output in pacemaker-dependent patients was caused by a proprietary algorithm designed to prevent T-wave sensing.
Hunyor I +5 more
europepmc +4 more sources
Postural change-dependent T-wave oversensing resulting in the administration of inappropriate shocks [PDF]
A 68-year-old man with dilated cardiomyopathy (left ventricular ejection fraction, 15%) and nonsustained ventricular tachycardia received an implantable cardioverter defibrillator. Even though his cardiac status had greatly improved 2 years later after β-
Yukoh Hirai +3 more
doaj +2 more sources
Introduction Extravascular implantable cardioverter‐defibrillator (EV‐ICD) is a novel therapy with a high‐voltage lead implanted in the substernal space.
Nobuhiro Nishii +5 more
doaj +2 more sources
Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure [PDF]
T-wave oversensing occurs when the counter starts giving dual beeps for every cardiac cycle instead of one. This usually happens when the monitoring lead displays a tall T wave, which is also sharp. R wave sensing algorithms of the devices do not sense T
Hassan Moladoust +2 more
doaj +4 more sources
Inappropriate pacing due to T-wave oversensing. [PDF]
Fernandez R +3 more
europepmc +3 more sources
Lead Integrity Alert Triggered by T-wave Oversensing. [PDF]
T-wave oversensing (TWOS) is a relatively common occurrence in pacemakers and defibrillators that can lead to pauses and inappropriate implantable cardioverter-defibrillator shocks. We present a case of TWOS that triggered the Lead Integrity Alert (Medtronic, Minneapolis, MN, USA) without evidence of actual lead failure.
Aboud AA, Clair WK.
europepmc +4 more sources
T-wave oversensing precipitating cardiogenic shock in a pacer-dependent patient. [PDF]
Adhikari S, Arisha MJ, Assal C.
europepmc +3 more sources
T-wave oversensing from postural changes: A rare cause of inappropriate shock from a subcutaneous defibrillator. [PDF]
Afzal MR +4 more
europepmc +2 more sources
Transient T-wave oversensing as a result of COVID-19: a case report. [PDF]
This case report explores the effects of myocarditis induced by coronavirus disease 2019 (COVID-19) on a cardiac implanted electronic device (CIED) and its ability to differentiate depolarisation and repolarisation. Through the modification of the device settings, inappropriate CIED discharges were prevented for the duration of the illness in this 76 ...
Khalil Z, Ward D, Ribiero C, French A.
europepmc +3 more sources